显微镜物镜加热器 OBJECTIVE HEATER 

Features

Temperature controlled heating band Easy installation around the objective Fits any objective on the market Compatible with the following OKOLAB Controllers: H301-T-UNIT-BL-PLUS UNO-T-H-PREMIXED UNO-T-H-CO2 H401-T-CONTROLLER Compatible with Inverted and Upright Microscopes Self CalibrationRoutines with BL controllers Voltage: 24 V Power: 13 W

    When should I use an Objective Heater You should use an objective heater, if you:

Perform Oil Immersion Imaging with a Top Stage Incubator Use dipping lens with an upright Top Stage Incubator Why should I use an Objective Heater? Because, if you don't, sample temperature will drop due to oil contact with the cold objective, as shown in the following graph.

 

Step 1: Objective not in contact with Petri 2; Objective Heater Off.

Step 2: Objective in oil contact with Petri 2; Objective Heater Off.

Step 3: Objective in oil contact with Petri 2; Objective Heater on, Offset value 5.3°C.

During the self calibration routine, the offset value of the objective heater is progressively increased, until T2 reaches the desired set point temperature of 37.0°C.

Figure 3 reports the profile of T1, T2, and of Objective Heater Temperature during the experimental campaign described above.